Fluorokine® MAP Multiplex Kits| Flourokine MAP Multiplex Mouse Cytokine Panel
Flourokine MAP Multiplex Mouse Cytokine Panel
Base Kit: LUM000 Mouse Fluorokine MAP Base Kit
Analyte Catalog Number Cell Culture Serum EDTA Plasma Sensitivity (pg/mL) High Standard Value* (pg/mL) Microparticle Region
GM-CSF LUM415B X X X 3.29 5500 11
IFN-gamma LUM485 X X X 7.60 16,500 75
IL-1 beta LUM401 X X X 7.82 35,400 06
IL-2 LUM402 X X X 6.30 9800 17
IL-4 LUM404 X 4.27 9800 21
IL-5 LUM405 X X X 4.18 8000 54
IL-6 LUM406 X X X 1.06 6600 32
IL-10 LUM417 X X X 1.28 5300 50
IL-12 p70 LUM419 X X X 22.42 23,000 66
IL-13 LUM413 X X X 42.0 40,800 36
IL-17 LUM421 X X X 10.8 9600 39
CCL2/JE LUM479 X X X 37.1 40,500 30
CXCL1/KC LUM453 X X X 2.69 9100 26
CXCL2/MIP-2 LUM452 X X X 10.1 4700 47
TNF-alpha LUM410 X X X 1.39 4700 79
VEGF LUM493 X X X 10.9 4400 43
*A standard curve must be generated each time an assay is run, utilizing values from the Standard Value Card included in the Base Kit.

Sample Preparation

Total prepared sample volume needed per well is 50 microliters.

Serum and plasma samples require a 4-fold dilution. For testing samples in duplicate, a suggested 4-fold dilution is 30 microliters of sample + 90 microliters of Calibrator Diluent RD6-40.

Cell culture supernates can be run undiluted with no matrix effects or interference. Maximum sample requirement per well is 50 microliters. Samples may be further diluted if necessary to bring sample data within the standard curve range.



RELATED INFORMATION
  • Bead Region Chart
  • Cytokines & Receptors
  • Fluorokine MAP Assay Principle for the Multiplex Kits
  • Multiplex Kits for the Luminex Platform
  • Recent Citations: Fluorokine MAP Multianalyte Profiling
  • Technical Notes: Multi-Cytokine Profiling