Human Immunotherapy Luminex® Performance Assay 24-plex Fixed Panel
R&D Systems | Catalog # LKTM010
Key Product Details
Assay Type
Format
Species
Analytes Detected
Not finding your analyte of interest? View our complete Luminex offering in the full version of our Luminex Assay Customization Tool. We also offer Custom Luminex Services.
Product Features
Magnetic Luminex High Performance Assays are flexible bead-based multiplex assays. They allow up to 50 user-defined target analytes to be simultaneously profiled using cell culture supernates, serum, or plasma samples. Some panels are also validated with human milk, saliva, and/or urine samples.
- Simultaneously profile Human Immunotherapy Assay Kits with a panel of related analytes in one sample
- Magnetic format allows for easier wash steps
- Undergoes similar validation testing as Quantikine® ELISA Kits
- Precise and Specific
- Requires a small sample volume (<50 µL)
- Run your assay in just 3-5 hours
- Mass-calibrated standards for consistent results with every new lot of material
- Enough reagents for one 96-well plate
- Quick turn-around time: Kits are typically shipped within 2-3 days of ordering*
*Please allow additional time for international orders.
Performance Validation
R&D Systems Luminex High Performance Assays undergo our most extensive validation testing. All Luminex High Performance Assay Panels are validated for use with serum and plasma samples. Some High Performance Assays are additionally validated for cell culture supernatants, milk, saliva, or urine, as indicated on the individual product data sheets. Luminex High Performance Assays are tested for sensitivity (three-quarters the low standard), intra-assay precision, inter-assay precision, and to ensure assay linearity for validated sample types. Recovery values for individual samples in validated sample types are also tested. Validation data for each analyte can be found in the product datasheet.
Base Kit Contents
- Premixed cocktail of antibody-coated Magnetic beads
- Premixed cocktail of biotinylated detection antibody
- Standard Cocktail(s)
- Bead Diluent
- Biotin Antibody Diluent
- Calibrator Diluent
- Wash Buffer
- Streptavidin-PE
- One flat-bottom 96-well Microplate
- Foil Plate Sealers (4)
- Mixing Bottle
- Standard Value Card**
**A standard curve must be generated each time an assay is run, utilizing values from the Certificate of Analysis provided with each Luminex Assay.
Precision
|
Intra-Assay Precision - is generated from the mean of the %CV's from 40 reportable results across two different concentrations of analytes in a single serum assay Inter-Assay Precision - is generated from the mean of the %CV's across two different concentrations of analytes across 25 different serum assays. |
||
| n=40 (20x2 levels) | n=25 | |
|---|---|---|
| Analyte | Intra-Assay (%CV) | Inter-Assay (%CV) |
| B7-H1/PD-L1 | 1.8% | 11.3% |
| CCL2/MCP-1 | 1.4% | 8.9% |
| CCL3/MIP-1alpha | 1.5% | 7.3% |
| CCL4/MIP-1beta | 3.0% | 9.6% |
| CD40 Ligand | 3.3% | 10.6% |
| CXCL10/IP-10 | 2.1% | 8.8% |
| CXCL8/IL-8 | 1.8% | 9.0% |
| GM-CSF | 3.1% | 10.3% |
| Granzyme B | 1.8% | 11.0% |
| IFN-alpha | 2.9% | 9.9% |
| IFN-gamma | 2.8% | 10.4% |
| IL-10 | 3.3% | 9.0% |
| IL-12 p70 | 2.3% | 10.5% |
| IL-13 | 1.6% | 9.7% |
| IL-15 | 2.4% | 9.9% |
| IL-17A | 2.5% | 11.2% |
| IL-1alpha | 3.0% | 10.4% |
| IL-1beta | 2.3% | 10.0% |
| IL-1ra | 2.1% | 10.3% |
| IL-2 | 2.2% | 9.2% |
| IL-33 | 2.4% | 12.3% |
| IL-4 | 3.2% | 9.7% |
| IL-6 | 2.6% | 9.8% |
| TNF-alpha | 2.2% | 8.7% |
Accuracy
|
Linearity - The data represents mean spiked or natural linearity in serum matrix samples. Samples with natural linearity are marked with an astrisk. Spike Recovery - The data represents mean percent recovery of spiked standards ranging from low, medium, and high concentration in serum matrix samples. |
||||
| Linearity | Spiked Recovery | |||
|---|---|---|---|---|
| Analyte | N | Average % | N | Average % |
|
*Natural linearity |
||||
| B7-H1/PD-L1 | 4 | 106% | 4 | 109% |
| CCL2/MCP-1 | 4* | 82% | High endogenous | |
| CCL3/MIP-1alpha | 4 | 77% | 4 | 139% |
| CCL4/MIP-1beta | 4 | 103% | 4 | 111% |
| CD40 Ligand | 4 | 113% | 4 | 100% |
| CXCL10/IP-10 | 3 | 110% | High endogenous | |
| CXCL8/IL-8 | 4 | 118% | 4 | 96% |
| GM-CSF | 4 | 96% | 4 | 110% |
| Granzyme B | 4 | 99% | 4 | 118% |
| IFN-alpha | 3 | 98% | 4 | 101% |
| IFN-gamma | 4 | 97% | 4 | 112% |
| IL-10 | 4 | 111% | 4 | 104% |
| IL-12 p70 | 4 | 106% | 4 | 108% |
| IL-13 | 4 | 112% | 4 | 103% |
| IL-15 | 3 | 88% | 3 | 89% |
| IL-17A | 4 | 110% | 4 | 106% |
| IL-1alpha | 4 | 143% | 4 | 80% |
| IL-1beta | 4 | 107% | 4 | 99% |
| IL-1ra | 4* | 100% | 4 | 145% |
| IL-2 | 4 | 105% | 4 | 102% |
| IL-33 | 4 | 93% | 4 | 110% |
| IL-4 | 3 | 127% | 3 | 77% |
| IL-6 | 3 | 124% | 3 | 90% |
| TNF-alpha | 4 | 117% | 4 | 69% |
Which Luminex Assay is Right for Me?
Assays for the Luminex platform are offered as High Performance Assays or Discovery Assays.
High Performance Assays are optimized and validated as complete panels with a focused set of analytes. You can choose from Fixed Panels, Premixed options, or Configurable User Mix formats depending on your needs.
Discovery Assays deliver maximum flexibility, enabling the creation of fully customized panels tailored to your research needs. With universal reagents validated for compatibility across all analytes in the menu, they support endless customization possibilities.
Additional Information About Luminex Assays
All R&D Systems Luminex Assays are designed for use with any Luminex instrument including MAGPIX, Luminex 200, FLEXMAP 3D, and the INTELLIFLEX.
Luminex assays use color-coded, magnetic microparticles coated with analyte-specific antibodies to enable multiplex quantitation. During the assay, standards and samples are added to the wells, where target analytes bind to the capture antibodies. A biotinylated detection antibody cocktail is then introduced, followed by a Streptavidin-PE conjugate that generates a fluorescent signal proportional to the amount of analyte present.
Each microparticle is identified by a unique signature and the signal intensity is measured using a Luminex instrument. These assays are designed to deliver reliable and reproducible results with validated antibody pairs and optimized protocols. They are well suited for applications that require sensitivity, accuracy, and confidence in your data while enabling efficient multiplexing to measure multiple targets simultaneously.
Scientific Data Images for Human Immunotherapy Luminex® Performance Assay 24-plex Fixed Panel
Human Immunotherapy Luminex Panel Standard Curves
Preparation and Storage
Shipping
Stability & Storage
Product Documents for Human Immunotherapy Luminex® Performance Assay 24-plex Fixed Panel
Product Specific Notices for Human Immunotherapy Luminex® Performance Assay 24-plex Fixed Panel
Luminex® is a registered trademark of Luminex Corporation.
For research use only
Citations for Human Immunotherapy Luminex® Performance Assay 24-plex Fixed Panel
Customer Reviews
There are currently no reviews for this product. Be the first to review Human Immunotherapy Luminex® Performance Assay 24-plex Fixed Panel and earn rewards!
Have you used Human Immunotherapy Luminex® Performance Assay 24-plex Fixed Panel?
Submit a review and receive an Amazon gift card!
$25/€18/£15/$25CAN/¥2500 Yen for a review with an image
$10/€7/£6/$10CAN/¥1110 Yen for a review without an image
Submit a review