Best Seller

Simple Plex AAV8 Cartridge

R&D Systems | Catalog # SPCKB-OT-007970

Simple Plex AAV8 assay kit contains cartridge, sample diluent SD19, and wash buffer. Part # starting with ST01 ships in 2-3 days.
R&D Systems
Loading...

Key Product Details

Assay Length

90 minutes

Sample Type & Volume Required Per Well*

Bioprocess Samples (25 µL)

Sensitivity

0 Not Assigned
0.1 106 capsids/mL (Bioprocess Samples)

Assay Range

0-0 Not Assigned
2.47-9428 106 capsids/mL (Bioprocess Samples)

Simple Plex AAV8 Cartridge Summary

Simple Plex assay for the detection of Adeno-Associated Virus type 8 (AAV8) intact capsids in bioprocess samples.This assay uses PROGEN's AAV8 (ADK8) antibody.

*The Sample Volume represented is based on the amount of sample incorporated into the reaction after taking into account the assay's minimum required dilution for a given matrix. Serial dilution may be necessary to achieve some of the final sample volumes represented.

Compatible Instruments

Ella Automated Immunoassay System

Ella Automated Immunoassay System

Efficiently analyze biomarkers with Ella's advanced technology, providing accurate, precise results in under 90 minutes. Quantify up to 8 targets per assay through a streamlined automated workflow. Choose from >350 highly validated Simple Plex assays or Simple Plex discovery assays.
Loading...

Product Specifications

Assay Type

Automated sandwich ELISA

Format

Microfluidic cartridge containing three Glass Nano Reactors (GNR)

Species

Viral

Specificity

This assay recognizes intact AAV8 capsids. The AAV8 antibody (ADK8) used in this assay cross-reacts with AAV1, AAV3, AAV7, and AAVrh10.

Cross-reactivity

< 0.5% cross-reactivity observed with available related molecules. < 50% cross-species reactivity observed with species tested.

Interference

No significant interference observed with available related molecules.

Precision

Intra-Assay Precision (Precision within an assay) Each control was tested 16 times in one assay.

Inter-Assay Precision (Precision between assays) Replicates of each control were tested in multiple assays performed by at least three technicians using two lots of reagents.

Bioprocess Samples

Intra-Assay Precision Inter-Assay Precision
Sample 1 2 1 2
n 16 16 47 45
Mean (10<sup>6</sup> capsids/mL) 23.1 1073 49.2 2202
Standard Deviation 0.658 55.4 4.51 99.3
CV% 2.9 5.2 9.2 4.5

Recovery for Simple Plex AAV8 Cartridge

Recovery was not necessary as AAV8 shows natural linearity data from a neat sample to a 1:16 dilution. 

Linearity

Samples containing and/or spiked with high concentrations of AAV8 were serially diluted with Sample Diluent to produce samples within the dynamic range of the assay.

Scientific Data Images for Simple Plex AAV8 Cartridge

AAV8 Simple Plex Assay Standard Curve

Preparation and Storage

Shipping

The product is shipped with polar packs. Upon receipt, store it immediately at the temperature recommended below.

Stability & Storage

Store the unopened product at 2 - 8 °C. Do not use past expiration date.

Background: AAV8

Long Name

Adeno-associated Virus 8

Additional AAV8 Products

Product Documents for Simple Plex AAV8 Cartridge

Loading product document information...

Loading product document information...

Certificate of Analysis

To download a Certificate of Analysis, please enter your Simple Plex catalog number and lot number in the search boxes below.

Product Specific Notices for Simple Plex AAV8 Cartridge

For research use only

Citations for Simple Plex AAV8 Cartridge

View all citations for the Simple Plex platform.

Please visit our Instrument Citation Database to search our collection of scientific articles that feature our instruments.

Customer Reviews for Simple Plex AAV8 Cartridge

There are currently no reviews for this product. Be the first to review Simple Plex AAV8 Cartridge and earn rewards!

Have you used Simple Plex AAV8 Cartridge?

Submit a review and receive an Amazon gift card!

$25/€18/£15/$25CAN/¥2500 Yen for a review with an image

$10/€7/£6/$10CAN/¥1110 Yen for a review without an image

Submit a review
Amazon Gift Card

FAQs for Simple Plex AAV8 Cartridge

Showing  1 - 5 of 6 FAQs Showing All
  • Q: Do the AAV assays recognize both natural and recombinant AAV?

    A: Yes, the AAV assays detect both natural and recombinant AAV. 

  • Q: Do the AAV assays recognize fully-assembled capsids?

    A: Yes, these assays detect fully-assembled capsids. Non-assembled protein or partially assembled capsids will not be detected.

  • Q: Do the Simple Plex AAV assays distinguish between full and empty capsids?

    A: No, the AAV assays do not distinguish between full and empty capsids. Complementary methods are required to distinguish full and empty capsids.

  • Q: Does the AAV8 assay recognize other AAV serotypes?

    A: This assay recognizes intact AAV8 capsids. The AAV8 antibody (ADK8) used in this assay cross-reacts with AAV1, AAV3, AAV7, and AAVrh10.

  • Q: What diluent is provided with the AAV assays?

    A: The AAV assays are supplied with SD19 concentrate. The concentrate should be diluted 1:5 in deionized or dilstilled water prior to use. 

  • Q: What is the minimum required dilution (MRD) for bio-process samples?

    A: All samples require a minimum 2-fold dilution for use in Simple Plex assays, but further dilution may be needed for optimal performace. An appropriate dilution factor for each process matrix should be determined experimentally by assessment of sample linearity and spike recovery. 

  • Q: Do the AAV assays recognize both natural and recombinant AAV?

    A: Yes, the AAV assays detect both natural and recombinant AAV. 

  • Q: Do the AAV assays recognize fully-assembled capsids?

    A: Yes, these assays detect fully-assembled capsids. Non-assembled protein or partially assembled capsids will not be detected.

  • Q: Do the Simple Plex AAV assays distinguish between full and empty capsids?

    A: No, the AAV assays do not distinguish between full and empty capsids. Complementary methods are required to distinguish full and empty capsids.

  • Q: Does the AAV8 assay recognize other AAV serotypes?

    A: This assay recognizes intact AAV8 capsids. The AAV8 antibody (ADK8) used in this assay cross-reacts with AAV1, AAV3, AAV7, and AAVrh10.

  • Q: What diluent is provided with the AAV assays?

    A: The AAV assays are supplied with SD19 concentrate. The concentrate should be diluted 1:5 in deionized or dilstilled water prior to use. 

  • Q: What is the minimum required dilution (MRD) for bio-process samples?

    A: All samples require a minimum 2-fold dilution for use in Simple Plex assays, but further dilution may be needed for optimal performace. An appropriate dilution factor for each process matrix should be determined experimentally by assessment of sample linearity and spike recovery. 

  • Q: Do the AAV assays recognize both natural and recombinant AAV?

    A: Yes, the AAV assays detect both natural and recombinant AAV. 

  • Q: Do the AAV assays recognize fully-assembled capsids?

    A: Yes, these assays detect fully-assembled capsids. Non-assembled protein or partially assembled capsids will not be detected.

  • Q: Do the Simple Plex AAV assays distinguish between full and empty capsids?

    A: No, the AAV assays do not distinguish between full and empty capsids. Complementary methods are required to distinguish full and empty capsids.

  • Q: Does the AAV8 assay recognize other AAV serotypes?

    A: This assay recognizes intact AAV8 capsids. The AAV8 antibody (ADK8) used in this assay cross-reacts with AAV1, AAV3, AAV7, and AAVrh10.

  • Q: What diluent is provided with the AAV assays?

    A: The AAV assays are supplied with SD19 concentrate. The concentrate should be diluted 1:5 in deionized or dilstilled water prior to use. 

  • Q: What is the minimum required dilution (MRD) for bio-process samples?

    A: All samples require a minimum 2-fold dilution for use in Simple Plex assays, but further dilution may be needed for optimal performace. An appropriate dilution factor for each process matrix should be determined experimentally by assessment of sample linearity and spike recovery. 

  • Q: Do the AAV assays recognize both natural and recombinant AAV?

    A: Yes, the AAV assays detect both natural and recombinant AAV. 

  • Q: Do the AAV assays recognize fully-assembled capsids?

    A: Yes, these assays detect fully-assembled capsids. Non-assembled protein or partially assembled capsids will not be detected.

  • Q: Do the Simple Plex AAV assays distinguish between full and empty capsids?

    A: No, the AAV assays do not distinguish between full and empty capsids. Complementary methods are required to distinguish full and empty capsids.

  • Q: Does the AAV8 assay recognize other AAV serotypes?

    A: This assay recognizes intact AAV8 capsids. The AAV8 antibody (ADK8) used in this assay cross-reacts with AAV1, AAV3, AAV7, and AAVrh10.

  • Q: What diluent is provided with the AAV assays?

    A: The AAV assays are supplied with SD19 concentrate. The concentrate should be diluted 1:5 in deionized or dilstilled water prior to use. 

  • Q: What is the minimum required dilution (MRD) for bio-process samples?

    A: All samples require a minimum 2-fold dilution for use in Simple Plex assays, but further dilution may be needed for optimal performace. An appropriate dilution factor for each process matrix should be determined experimentally by assessment of sample linearity and spike recovery. 

  • Q: Do the AAV assays recognize both natural and recombinant AAV?

    A: Yes, the AAV assays detect both natural and recombinant AAV. 

  • Q: Do the AAV assays recognize fully-assembled capsids?

    A: Yes, these assays detect fully-assembled capsids. Non-assembled protein or partially assembled capsids will not be detected.

  • Q: Do the Simple Plex AAV assays distinguish between full and empty capsids?

    A: No, the AAV assays do not distinguish between full and empty capsids. Complementary methods are required to distinguish full and empty capsids.

  • Q: Does the AAV8 assay recognize other AAV serotypes?

    A: This assay recognizes intact AAV8 capsids. The AAV8 antibody (ADK8) used in this assay cross-reacts with AAV1, AAV3, AAV7, and AAVrh10.

  • Q: What diluent is provided with the AAV assays?

    A: The AAV assays are supplied with SD19 concentrate. The concentrate should be diluted 1:5 in deionized or dilstilled water prior to use. 

  • Q: What is the minimum required dilution (MRD) for bio-process samples?

    A: All samples require a minimum 2-fold dilution for use in Simple Plex assays, but further dilution may be needed for optimal performace. An appropriate dilution factor for each process matrix should be determined experimentally by assessment of sample linearity and spike recovery. 

  • Q: Do the AAV assays recognize both natural and recombinant AAV?

    A: Yes, the AAV assays detect both natural and recombinant AAV. 

  • Q: Do the AAV assays recognize fully-assembled capsids?

    A: Yes, these assays detect fully-assembled capsids. Non-assembled protein or partially assembled capsids will not be detected.

  • Q: Do the Simple Plex AAV assays distinguish between full and empty capsids?

    A: No, the AAV assays do not distinguish between full and empty capsids. Complementary methods are required to distinguish full and empty capsids.

  • Q: Does the AAV8 assay recognize other AAV serotypes?

    A: This assay recognizes intact AAV8 capsids. The AAV8 antibody (ADK8) used in this assay cross-reacts with AAV1, AAV3, AAV7, and AAVrh10.

  • Q: What diluent is provided with the AAV assays?

    A: The AAV assays are supplied with SD19 concentrate. The concentrate should be diluted 1:5 in deionized or dilstilled water prior to use. 

  • Q: What is the minimum required dilution (MRD) for bio-process samples?

    A: All samples require a minimum 2-fold dilution for use in Simple Plex assays, but further dilution may be needed for optimal performace. An appropriate dilution factor for each process matrix should be determined experimentally by assessment of sample linearity and spike recovery. 

Showing  1 - 5 of 6 FAQs Showing All