Precision
Bioprocess Samples
Intra-Assay Precision | Inter-Assay Precision | |||
---|---|---|---|---|
Sample | 1 | 2 | 1 | 2 |
n | 8 | 8 | 81 | 79 |
Mean (ng/mL) | 3.16 | 147 | 3.36 | 141 |
Standard Deviation | 0.38 | 7.3 | 0.48 | 14.9 |
CV% | 12 | 5 | 14.3 | 10.6 |
Recovery
Recovery at two different spiked concentrations within the range of the assay was evaluated.
Sample Type | Average % Recovery | Range % |
---|---|---|
Bioprocess Samples (n=12) | 111 | 100-123 |
Linearity
Scientific Data
Product Datasheets
Preparation and Storage
FAQs
-
What is the minimum required dilution (MRD) for bio-process samples?
All samples require a minimum 2-fold dilution for use in Simple Plex assays, but further dilution may be needed for optimal performace. An appropriate dilution factor for each process matrix should be determined experimentally by assessment of sample linearity and spike recovery.