Simple Plex Human CXCL10/IP-10 Cartridge
Simple Plex Human CXCL10/IP-10 Cartridge Summary
*The Sample Volume represented is based on the amount of sample incorporated into the reaction after taking into account the assay’s minimum required dilution for a given matrix. Serial dilution may be necessary to achieve some of the final sample volumes represented.
Product Summary
Precision
Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Urine, Cerebrospinal Fluid
| Intra-Assay Precision | Inter-Assay Precision | |||
|---|---|---|---|---|
| Sample | 1 | 2 | 1 | 2 |
| n | 16 | 16 | 14 | 12 |
| Mean (pg/mL) | 5.56 | 283 | 5.49 | 297 |
| Standard Deviation | 0.347 | 14.3 | 0.212 | 11.7 |
| CV% | 6.1 | 5 | 3.9 | 3.9 |
Recovery
Recovery at three different spiked concentrations within the range of the assay was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| EDTA Plasma (n=4) | 87 | 74-105 |
| Heparin Plasma (n=4) | 92 | 85-103 |
| Serum (n=4) | 93 | 80-107 |
| Urine (n=8) | 100 | 90-115 |
Linearity
Scientific Data
Product Datasheets
Preparation and Storage
Background: CXCL10/IP-10/CRG-2
IP-10 was originally identified as an IFN-gamma-inducible gene in monocytes, fibroblasts and endothelial cells. The mouse homolog of human IP-10, CRG-2, shares approximately 67% amino acid sequence identity with human IP-10. The amino acid sequence of IP-10 identified the protein as a member of the CXC chemokine subfamily.
FAQs
No product specific FAQs exist for this product, however you may
View all Simple Plex Assays FAQsReviews for Simple Plex Human CXCL10/IP-10 Cartridge
There are currently no reviews for this product. Be the first to review Simple Plex Human CXCL10/IP-10 Cartridge and earn rewards!
Have you used Simple Plex Human CXCL10/IP-10 Cartridge?
Submit a review and receive an Amazon gift card.
$25/€18/£15/$25CAN/¥75 Yuan/¥2500 Yen for a review with an image
$10/€7/£6/$10 CAD/¥70 Yuan/¥1110 Yen for a review without an image