Researchers can encounter a variety of problems when running an ELISA. Explore this page to understand the potential source of your problem and the steps we recommend for correcting the issue to help set you on a path for running a successful ELISA. For further assistance, please contact our technical service department.
Problem: High Background
| Possible Source | Test or Action |
|---|
| Insufficient washing | See washing procedure on page 4 of the ELISA Development Guide |
| Increase number of washes Add a 30 second soak step inbetween washes |
Back to Top
Problem: No signal when a signal is expected
| Possible Source | Test or Action |
|---|
| Reagents added in incorrect order, or incorrectly prepared | Repeat assay Check calculations and make new buffers, standards, etc.
Review protocol |
| Contamination of HRP with | Use fresh reagents |
| Not enough antibody used | Increase concentration |
| Standard has gone bad (if there is a signal in the sample wells) | Check that standard was handled according to directions.
Use new vial. |
| Buffer containing FCS used to reconstitute antibodies | Requalify your reagents of choice |
| Capture antibody did not bind to plate | Use an ELISA plate (not a tissue culture plate)
Dilute in PBS without additional protein |
| Buffers contaminated | Make fresh buffers |
Back to Top
Problem: Too much signal - whole plate turned uniformly blue
| Possible Source | Test or Action |
|---|
| Insufficient washing/washing step skipped - unbound peroxidase remaining | See washing procedure on page 4 of the ELISA Development Guide |
| Substrate Solution mixed too early and turned blue | Substrate Solution should be mixed and used immediately |
| Too much streptavidin-HRP | Check dilution, titrate if necessary |
| Plate sealers or reagent reservoirs reused, resulting in presence of residual HRP. This will turn the TMB blue non-specifically | Use fresh plate sealer and reagent reservoir for each step |
| Buffers contaminated with metals or HRP | Make fresh buffers |
Back to Top
Problem: Standard curve achieved but poor discrimination between points (low or flat curve)
| Possible Source | Test or Action |
|---|
| Not enough streptavidin-HRP | Check dilution, titrate if necessary |
| Capture antibody did not bind well to plate | Use an ELISA plate (not a tissue culture plate) Dilute in PBS without additional protein |
| Not enough detection antibody | Check dilution, titrate if necessary |
| Plate not developed long enough | Increase Substrate Solution incubation time
Use recommended brand of Substrate Solution |
| Incorrect procedure | Go back to General ELISA Protocol; eliminate modifications, if any |
| Improper calculation of standard curve dilutions | Check calculations, make new standard curve |
Back to Top
Problem: Poor Duplicates
| Possible Source | Test or Action |
|---|
| Insufficient washing | See washing procedures on page 4 of the ELISA Development Guide
If using an automatic plate washer, check that all ports are clean and free of obstructions, add a 30 second soak step and rotate plate halfway through the wash |
| Uneven plate coating due to procedural error or poor plate quality (can bind unevenly) | Dilute in PBS without additional protein
Check coating and blocking volumes, times and method of reagent addition. Check plate used
Use an ELISA plate (not a tissue culture plate) |
| Plate sealer reused | Use a fresh plate sealer for each step |
| No plate sealers used | Use plate sealers |
| Buffers contaminated | Make fresh buffers |
Back to Top
Problem: Poor assay to assay reproducibility
| Possible Source | Test or Action |
|---|
| Insufficient washing | See washing procedure on page 4 of the ELISA Development Guide
If using an automatic plate washer, check that all ports are clean and free of obstructions |
| Variations in incubation temperature | Adhere to recommended incubation temperature
Avoid incubating plates in areas where enviromental conditions vary |
| Variations in protocol | Adhere to the same protocol from run to run |
| Plate sealer reused, resulting in presence of residual HRP which will turn the TMB blue | Use fresh plate sealer for each step |
| Improper calculation of standard curve dilutions | Check calculations, make new standard curve
Use internal controls |
| Buffers contaminated | Make fresh buffers |
Back to Top
Problem: No signal when a signal is expected, but standard curve looks fine
| Possible Source | Test or Action |
|---|
| No cytokine in sample | Use internal controls
Repeat experiment, reconsider experimental parameters |
| Sample matrix is masking detection | Dilute samples at least 1:2 in appropriate diluent, or preferably, do a series of dilutions to look at recovery |
Back to Top
Problem: Samples are reading too high, but standard curve looks fine
| Possible Source | Test or Action |
|---|
| Samples contain cytokine levels above assay range | Dilute samples and run again |
Back to Top
Problem: Very low readings across the plate
| Possible Source | Test or Action |
|---|
| Incorrect wavelengths | Check filters/reader |
| Insufficient development time | Increase development time |
| Coated plates are old and have gone bad | Coat new plates |
| Capture antibody did not bind to the plate | Use an ELISA plate (not a tissue culture plate)
Dilute in PBS without additional protein |
| Buffer containing FCS used to reconstitute antibodies | Requalify your reagents of choice |
Back to Top
Problem: Green color develops upon addition of stop solution when using streptavidin-HRP
| Possible Source | Test or Action |
|---|
| Reagents not mixed well enough in wells | Tap plate |
Back to Top
Problem: Edge Effects
| Possible Source | Test or Action |
|---|
| Uneven temperatures around work surface | Avoid incubating plates in areas where environmental conditions vary |
| Use plate sealers |
Back to Top
Problem: Drift
| Possible Source | Test or Action |
|---|
| Interrupted assay set-up | Assay set-up should be continuous - have all standards and samples prepared appropriately before commencement of the assay |
| Reagents not at room temperature | Ensure that all reagents are at room temperature before pipetting into the wells unless otherwise instructed in the antibody inserts |
ELISA Kits, Controls and Services
ELISA Protocols, Assay Principles and Guides
ELISA Kit 101: Introduction to the ELISA Technique
ELISA Articles and Additional Resources
Why choose R&D Systems ELISA Kits?
Blogs
Literature
Handbooks
Videos
Back to Top
ELISA Protocols
Related Information
DuoSet ELISA Development Systems
DuoSet IC (IntraCellular) Assay Development Systems
ELISA Development Guide (PDF 152K)
ELISA Guide (PDF)