Human VEGF Quantikine ELISA Kit
R&D Systems | Catalog # DVE00
Key Product Details
Assay Length
Sample Type & Volume Required Per Well
Sensitivity
9 pg/mL (Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma)
Assay Range
31.3-2000 pg/mL (Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma)
Product Summary for Human VEGF Quantikine ELISA Kit
Product Specifications
Assay Type
Format
Measurement
Detection Method
Conjugate
Species
Specificity
Cross-reactivity
Interference
Sample Values
| Sample Type | Mean of detectable (pg/mL) | % Detectable | Range (pg/mL) |
| Serum (n=37) | 220 | 100 | 62-707 |
| EDTA plasma (n=37) | 61 | 24 | ND-115 |
| Heparin plasma (n=37) | 41 | 22 | ND-55 |
| Citrate plasma (n=37) | ___ | 0 | ND |
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in forty separate assays to assess inter-assay precision. Assays were performed by at least three technicians using two lots of components.
Cell Culture Supernates
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean (pg/mL) | 29.1 | 123 | 531 | 32.8 | 128 | 495 |
| Standard Deviation | 1.9 | 5.0 | 18.4 | 2.8 | 6.4 | 33.0 |
| CV% | 6.5 | 4.1 | 3.5 | 8.5 | 5.0 | 6.7 |
Citrate Plasma, EDTA Plasma, Heparin Plasma, Serum
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 40 | 40 | 40 |
| Mean (pg/mL) | 53.7 | 235 | 910 | 64.5 | 250 | 1003 |
| Standard Deviation | 3.6 | 10.6 | 46.2 | 5.7 | 17.4 | 61.7 |
| CV% | 6.7 | 4.5 | 5.1 | 8.8 | 7.0 | 6.2 |
Recovery for Human VEGF Quantikine ELISA Kit
The recovery of human VEGF spiked to three different levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=5) | 102 | 95-111 |
| Citrate Plasma (n=5) | 100 | 88-113 |
| EDTA Plasma (n=5) | 97 | 82-113 |
| Heparin Plasma (n=5) | 93 | 82-102 |
| Serum (n=5) | 102 | 92-115 |
Linearity
To assess linearity of the assay, samples were spiked with high concentrations of human VEGF and diluted with the appropriate calibrator diluent to produce samples with values within the dynamic range of the assay.
Scientific Data Images for Human VEGF Quantikine ELISA Kit
Human VEGF ELISA Cell Culture Supernate Standard Curve
Human VEGF ELISA Serum/Plasma Standard Curve
Preparation and Storage
Shipping
Stability & Storage
Background: VEGF
Long Name
Alternate Names
Entrez Gene IDs
Gene Symbol
Additional VEGF Products
Product Documents for Human VEGF Quantikine ELISA Kit
Certificate of Analysis
To download a Certificate of Analysis, please enter a lot or batch number in the search box below.
Note: Certificate of Analysis not available for kit components.
Product Specific Notices for Human VEGF Quantikine ELISA Kit
For research use only
⚠ WARNING: This product can expose you to chemicals including N,N-Dimethylforamide, which is known to the State of California to cause cancer. For more information, go to www.P65Warnings.ca.gov.Citations for Human VEGF Quantikine ELISA Kit
Customer Reviews for Human VEGF Quantikine ELISA Kit (29)
Have you used Human VEGF Quantikine ELISA Kit?
Submit a review and receive an Amazon gift card!
$25/€18/£15/$25CAN/¥2500 Yen for a review with an image
$10/€7/£6/$10CAN/¥1110 Yen for a review without an image
Submit a review
Customer Images
-
Sample Tested: Serum and PlasmaVerified Customer | Posted 11/10/2023
-
Sample Tested: Serum and Lung cancer tissueVerified Customer | Posted 10/31/2023We ordered a large batch of reagent kits together to study the interaction between blood vessel formation and the microenvironment. The results are evident, and the reagent kit testing is very sensitive.
-
Sample Tested: Human fibroblastVerified Customer | Posted 10/11/2022Human VEGF was measured in human HUVEC Cells after stimulation with IFN-g and IL-1b. An increased production was significantly obtained, compared to unstimulated cells.
-
Sample Tested: Serum-free Cell Culture SupernatesVerified Customer | Posted 11/24/2020This ELISA kit performed as expected. Very low % CV's
-
Sample Tested: HeLa cellsVerified Customer | Posted 08/26/2020
-
Sample Tested: Serum-free Cell Culture MediaVerified Customer | Posted 01/07/2020
-
Sample Tested: Cell culture supernatantVerified Customer | Posted 12/09/2019
-
Sample Tested: in vitro cell cultureVerified Customer | Posted 10/19/2019
-
Sample Tested: 4T1.2 mouse mammary carcinoma cellsVerified Customer | Posted 10/17/2019
-
Sample Tested: Bone ExtractsVerified Customer | Posted 10/10/2019
-
Sample Tested: in vitro cell lysisVerified Customer | Posted 08/13/2019
-
Sample Tested: HUVEC human umbilical vein endothelial cellsVerified Customer | Posted 04/30/2019
-
Sample Tested: Brain Microvascular Endothelial CellsVerified Customer | Posted 03/21/2019
-
Sample Tested: Cell Culture MediaVerified Customer | Posted 11/09/2018Just follow the protocol. Very easy.
-
Sample Tested: SerumVerified Customer | Posted 06/12/2018
-
Sample Tested: B16-F10 mouse melanoma cell lineVerified Customer | Posted 04/18/2018
-
Sample Tested: SerumVerified Customer | Posted 04/05/2018
-
Sample Tested: SerumVerified Customer | Posted 03/30/2018
-
Sample Tested: In vitro biochemical assayVerified Customer | Posted 03/22/2018
-
Sample Tested: SerumVerified Customer | Posted 01/12/2018
-
Sample Tested: Human cell conditioned medium, A375 human melanoma cell line and SK-Mel-28 human malignant melanoma cell lineVerified Customer | Posted 10/06/2017I have used this kit for a very large part of my PhD project on >20 cell lines. It worked every time, giving reproducible and reliable results with great sensibility. Highly recommend this.
-
Sample Tested: Cell Culture MediaVerified Customer | Posted 09/27/2017
-
Sample Tested: Human skeletal muscleVerified Customer | Posted 04/21/2017Followed instructions for serum/plasma. Ran samples full strength
-
Sample Tested: EDTA PlasmaVerified Customer | Posted 01/23/2017This VEGF ELISA performs very well and very reproducible. We have used different lot#s, all perform very equally. Well calibrated Standard Protein si provided, curve is almost identical to reference Standard (NIBSC). We usually extend the Standard curve by an additional low calibrator Point (additional 1:2 Dilution of S7), because the Assay still performes well at the low Assay range. Human EDTA-Plasma samples are usually quantifiable and above the detection limit. In general human blood samples matrix should be selected carefully to avoid effects from VEGF release during coagulation.
-
Sample Tested: Hematopoietic Stem CellsVerified Customer | Posted 12/14/2016In our lab, we developed serum-free culture media to expand human hematopoietic stem cells and human endothelial progenitor cells in vitro. We also tried to cocluture both cells. After culture, Human VEGF Quantikine ELISA Kit were used to check the cytokine, VEGF secretion.
-
Sample Tested: iPSC-derived neural cellsVerified Customer | Posted 12/14/2016I used this ELISA Kit to measure VEGF secretion in iPSC-derived astrocytes, brain microvascular endothelial cells and neurons following exposure to hypoxic/ischemic injury in vitro. The kit was easy to use, well detailed and provided enough reagents to successfully run the assay.
-
Sample Tested: Adipose tissueVerified Customer | Posted 06/20/2016
-
Sample Tested: IMR-32 human neuroblastoma cell lineVerified Customer | Posted 06/17/2016
-
Sample Tested: Bone PuttyVerified Customer | Posted 05/09/2016
There are no reviews that match your criteria.
Protocols
View specific protocols for Human VEGF Quantikine ELISA Kit (DVE00):
Refer to the product datasheet for the complete assay procedure.
Bring all reagents and samples to room temperature before use. It is recommended that all samples, standards, and controls be assayed in duplicate.
- Prepare all reagents, standard dilutions, and samples as directed in the product insert.
- Remove excess microplate strips from the plate frame, return them to the foil pouch containing the desiccant pack, and reseal.
-
For Serum & Plasma Samples: Add 100 µL of Assay Diluent to each well.
For Cell Culture Supernate Samples: Add 50 µL Assay Diluent to each well. -
For Serum & Plasma Samples: Add 100 µL of Standard, control, or sample to each well. Cover with a plate sealer, and incubate at room temperature for 2 hours.
For Cell Culture Supernate Samples: Add 200 µL of Standard, control, or sample to each well. Cover with a plate sealer, and incubate at room temperature for 2 hours. - Aspirate each well and wash, repeating the process twice for a total of 3 washes.
- Add 200 µL of Conjugate to each well. Cover with a new plate sealer, and incubate at room temperature for 2 hours.
- Aspirate and wash 3 times.
- Add 200 µL Substrate Solution to each well.
-
For Serum & Plasma Samples: Incubate at room temperature for 25 minutes. PROTECT FROM LIGHT.
For Cell Culture Supernate Samples: Incubate at room temperature for 20 minutes.
PROTECT FROM LIGHT. -
For Serum & Plasma Samples: Incubate at room temperature for 25 minutes. PROTECT FROM LIGHT.
For Cell Culture Supernate Samples: Incubate at room temperature for 20 minutes.
PROTECT FROM LIGHT.
Add Assay Diluent
Add Standard, Control, or Sample
200 µL Conjugate
200 µL Substrate Solution
50 µL Stop Solution
Find general support by application which include: protocols, troubleshooting, illustrated assays, videos and webinars.
- ELISA Sample Preparation & Collection Guide
- ELISA Troubleshooting Guide
- How to Run an R&D Systems DuoSet ELISA
- How to Run an R&D Systems Quantikine ELISA
- How to Run an R&D Systems Quantikine™ QuicKit™ ELISA
- Quantikine HS ELISA Kit Assay Principle, Alkaline Phosphatase
- Quantikine HS ELISA Kit Principle, Streptavidin-HRP Polymer
- Sandwich ELISA (Colorimetric) – Biotin/Streptavidin Detection Protocol
- Sandwich ELISA (Colorimetric) – Direct Detection Protocol
- Troubleshooting Guide: ELISA
- View all Protocols, Troubleshooting, Illustrated assays and Webinars
FAQs for Human VEGF Quantikine ELISA Kit
-
Q: Does the Human VEGF Quantikine® ELISA kit (Catalog # DVE00) detect human VEGF121?
A: Yes, the kit detects VEGF121. In specificity testing, 100% cross-reactivity was observed with recombinant human VEGF121.
-
Q: Does the Human VEGF Quantikine® ELISA kit (Catalog # DVE00) detect mouse VEGF?
A: No, the kit does not detect mouse VEGF. Specificity testing with recombinant mouse VEGF 120 and VEGF 164 showed no cross-reactivity. For the detection of mouse VEGF, please refer to the Mouse VEGF Quantikine® ELISA kit (Catalog # MMV00).
-
Q: Does the Human VEGF Quantikine® ELISA kit (Catalog # DVE00) detect porcine samples?
A: Our lab has not tested porcine samples in the kit, however many customers have published use of the kit with porcine samples. Links to these references can be found in the citations tab by selecting for Porcine in the Species filter.
-
Q: How long can samples (cell cultures, serum, and plasma) be stored and still be stable for measuring VEGF using the DVE00 Human VEGF Quantikine® ELISA Kit?
A: R&D Systems does not evaluate sample stability. Variations in sample collection, processing, and storage may affect stability of samples. It is recommend to assay sample immediately after collection when possible, or aliquot into single use volumes and store samples frozen. Avoid repeat freeze-thaw cycles with the stored samples to avoid protein degradation.
-
Q: Is the heterodimer VEGF165/PlGF detected by DVE00?
A: VEGF165/PlGF is not detected by this kit. Recombinant human VEGF165/PlGF was prepared at 50 ng/mL and assayed for cross-reactivity, with negative results.
-
Q: Is there a protocol for using cell lysates with the Human VEGF Quantikine® ELISA Kit (Catalog # DVE00)?
A: R&D Systems does not have a validated cell lysate protocol available for this kit. However, many customers have published use of the kit with cell lysates. Links to these references can be found in the citations tab. Generally, the use of detergents in the lysis buffer should be avoided when preparing cell samples. A better option is to lyse the cells via freeze/thaw cycles or by sonication.
-
Q: Does the Human VEGF Quantikine® ELISA kit (Catalog # DVE00) detect human VEGF121?
A: Yes, the kit detects VEGF121. In specificity testing, 100% cross-reactivity was observed with recombinant human VEGF121.
-
Q: Does the Human VEGF Quantikine® ELISA kit (Catalog # DVE00) detect mouse VEGF?
A: No, the kit does not detect mouse VEGF. Specificity testing with recombinant mouse VEGF 120 and VEGF 164 showed no cross-reactivity. For the detection of mouse VEGF, please refer to the Mouse VEGF Quantikine® ELISA kit (Catalog # MMV00).
-
Q: Does the Human VEGF Quantikine® ELISA kit (Catalog # DVE00) detect porcine samples?
A: Our lab has not tested porcine samples in the kit, however many customers have published use of the kit with porcine samples. Links to these references can be found in the citations tab by selecting for Porcine in the Species filter.
-
Q: How long can samples (cell cultures, serum, and plasma) be stored and still be stable for measuring VEGF using the DVE00 Human VEGF Quantikine® ELISA Kit?
A: R&D Systems does not evaluate sample stability. Variations in sample collection, processing, and storage may affect stability of samples. It is recommend to assay sample immediately after collection when possible, or aliquot into single use volumes and store samples frozen. Avoid repeat freeze-thaw cycles with the stored samples to avoid protein degradation.
-
Q: Is the heterodimer VEGF165/PlGF detected by DVE00?
A: VEGF165/PlGF is not detected by this kit. Recombinant human VEGF165/PlGF was prepared at 50 ng/mL and assayed for cross-reactivity, with negative results.
-
Q: Is there a protocol for using cell lysates with the Human VEGF Quantikine® ELISA Kit (Catalog # DVE00)?
A: R&D Systems does not have a validated cell lysate protocol available for this kit. However, many customers have published use of the kit with cell lysates. Links to these references can be found in the citations tab. Generally, the use of detergents in the lysis buffer should be avoided when preparing cell samples. A better option is to lyse the cells via freeze/thaw cycles or by sonication.
-
Q: Does the Human VEGF Quantikine® ELISA kit (Catalog # DVE00) detect human VEGF121?
A: Yes, the kit detects VEGF121. In specificity testing, 100% cross-reactivity was observed with recombinant human VEGF121.
-
Q: Does the Human VEGF Quantikine® ELISA kit (Catalog # DVE00) detect mouse VEGF?
A: No, the kit does not detect mouse VEGF. Specificity testing with recombinant mouse VEGF 120 and VEGF 164 showed no cross-reactivity. For the detection of mouse VEGF, please refer to the Mouse VEGF Quantikine® ELISA kit (Catalog # MMV00).
-
Q: Does the Human VEGF Quantikine® ELISA kit (Catalog # DVE00) detect porcine samples?
A: Our lab has not tested porcine samples in the kit, however many customers have published use of the kit with porcine samples. Links to these references can be found in the citations tab by selecting for Porcine in the Species filter.
-
Q: How long can samples (cell cultures, serum, and plasma) be stored and still be stable for measuring VEGF using the DVE00 Human VEGF Quantikine® ELISA Kit?
A: R&D Systems does not evaluate sample stability. Variations in sample collection, processing, and storage may affect stability of samples. It is recommend to assay sample immediately after collection when possible, or aliquot into single use volumes and store samples frozen. Avoid repeat freeze-thaw cycles with the stored samples to avoid protein degradation.
-
Q: Is the heterodimer VEGF165/PlGF detected by DVE00?
A: VEGF165/PlGF is not detected by this kit. Recombinant human VEGF165/PlGF was prepared at 50 ng/mL and assayed for cross-reactivity, with negative results.
-
Q: Is there a protocol for using cell lysates with the Human VEGF Quantikine® ELISA Kit (Catalog # DVE00)?
A: R&D Systems does not have a validated cell lysate protocol available for this kit. However, many customers have published use of the kit with cell lysates. Links to these references can be found in the citations tab. Generally, the use of detergents in the lysis buffer should be avoided when preparing cell samples. A better option is to lyse the cells via freeze/thaw cycles or by sonication.
-
Q: Does the Human VEGF Quantikine® ELISA kit (Catalog # DVE00) detect human VEGF121?
A: Yes, the kit detects VEGF121. In specificity testing, 100% cross-reactivity was observed with recombinant human VEGF121.
-
Q: Does the Human VEGF Quantikine® ELISA kit (Catalog # DVE00) detect mouse VEGF?
A: No, the kit does not detect mouse VEGF. Specificity testing with recombinant mouse VEGF 120 and VEGF 164 showed no cross-reactivity. For the detection of mouse VEGF, please refer to the Mouse VEGF Quantikine® ELISA kit (Catalog # MMV00).
-
Q: Does the Human VEGF Quantikine® ELISA kit (Catalog # DVE00) detect porcine samples?
A: Our lab has not tested porcine samples in the kit, however many customers have published use of the kit with porcine samples. Links to these references can be found in the citations tab by selecting for Porcine in the Species filter.
-
Q: How long can samples (cell cultures, serum, and plasma) be stored and still be stable for measuring VEGF using the DVE00 Human VEGF Quantikine® ELISA Kit?
A: R&D Systems does not evaluate sample stability. Variations in sample collection, processing, and storage may affect stability of samples. It is recommend to assay sample immediately after collection when possible, or aliquot into single use volumes and store samples frozen. Avoid repeat freeze-thaw cycles with the stored samples to avoid protein degradation.
-
Q: Is the heterodimer VEGF165/PlGF detected by DVE00?
A: VEGF165/PlGF is not detected by this kit. Recombinant human VEGF165/PlGF was prepared at 50 ng/mL and assayed for cross-reactivity, with negative results.
-
Q: Is there a protocol for using cell lysates with the Human VEGF Quantikine® ELISA Kit (Catalog # DVE00)?
A: R&D Systems does not have a validated cell lysate protocol available for this kit. However, many customers have published use of the kit with cell lysates. Links to these references can be found in the citations tab. Generally, the use of detergents in the lysis buffer should be avoided when preparing cell samples. A better option is to lyse the cells via freeze/thaw cycles or by sonication.
-
Q: Does the Human VEGF Quantikine® ELISA kit (Catalog # DVE00) detect human VEGF121?
A: Yes, the kit detects VEGF121. In specificity testing, 100% cross-reactivity was observed with recombinant human VEGF121.
-
Q: Does the Human VEGF Quantikine® ELISA kit (Catalog # DVE00) detect mouse VEGF?
A: No, the kit does not detect mouse VEGF. Specificity testing with recombinant mouse VEGF 120 and VEGF 164 showed no cross-reactivity. For the detection of mouse VEGF, please refer to the Mouse VEGF Quantikine® ELISA kit (Catalog # MMV00).
-
Q: Does the Human VEGF Quantikine® ELISA kit (Catalog # DVE00) detect porcine samples?
A: Our lab has not tested porcine samples in the kit, however many customers have published use of the kit with porcine samples. Links to these references can be found in the citations tab by selecting for Porcine in the Species filter.
-
Q: How long can samples (cell cultures, serum, and plasma) be stored and still be stable for measuring VEGF using the DVE00 Human VEGF Quantikine® ELISA Kit?
A: R&D Systems does not evaluate sample stability. Variations in sample collection, processing, and storage may affect stability of samples. It is recommend to assay sample immediately after collection when possible, or aliquot into single use volumes and store samples frozen. Avoid repeat freeze-thaw cycles with the stored samples to avoid protein degradation.
-
Q: Is the heterodimer VEGF165/PlGF detected by DVE00?
A: VEGF165/PlGF is not detected by this kit. Recombinant human VEGF165/PlGF was prepared at 50 ng/mL and assayed for cross-reactivity, with negative results.
-
Q: Is there a protocol for using cell lysates with the Human VEGF Quantikine® ELISA Kit (Catalog # DVE00)?
A: R&D Systems does not have a validated cell lysate protocol available for this kit. However, many customers have published use of the kit with cell lysates. Links to these references can be found in the citations tab. Generally, the use of detergents in the lysis buffer should be avoided when preparing cell samples. A better option is to lyse the cells via freeze/thaw cycles or by sonication.
-
Q: Does the Human VEGF Quantikine® ELISA kit (Catalog # DVE00) detect human VEGF121?
A: Yes, the kit detects VEGF121. In specificity testing, 100% cross-reactivity was observed with recombinant human VEGF121.
-
Q: Does the Human VEGF Quantikine® ELISA kit (Catalog # DVE00) detect mouse VEGF?
A: No, the kit does not detect mouse VEGF. Specificity testing with recombinant mouse VEGF 120 and VEGF 164 showed no cross-reactivity. For the detection of mouse VEGF, please refer to the Mouse VEGF Quantikine® ELISA kit (Catalog # MMV00).
-
Q: Does the Human VEGF Quantikine® ELISA kit (Catalog # DVE00) detect porcine samples?
A: Our lab has not tested porcine samples in the kit, however many customers have published use of the kit with porcine samples. Links to these references can be found in the citations tab by selecting for Porcine in the Species filter.
-
Q: How long can samples (cell cultures, serum, and plasma) be stored and still be stable for measuring VEGF using the DVE00 Human VEGF Quantikine® ELISA Kit?
A: R&D Systems does not evaluate sample stability. Variations in sample collection, processing, and storage may affect stability of samples. It is recommend to assay sample immediately after collection when possible, or aliquot into single use volumes and store samples frozen. Avoid repeat freeze-thaw cycles with the stored samples to avoid protein degradation.
-
Q: Is the heterodimer VEGF165/PlGF detected by DVE00?
A: VEGF165/PlGF is not detected by this kit. Recombinant human VEGF165/PlGF was prepared at 50 ng/mL and assayed for cross-reactivity, with negative results.
-
Q: Is there a protocol for using cell lysates with the Human VEGF Quantikine® ELISA Kit (Catalog # DVE00)?
A: R&D Systems does not have a validated cell lysate protocol available for this kit. However, many customers have published use of the kit with cell lysates. Links to these references can be found in the citations tab. Generally, the use of detergents in the lysis buffer should be avoided when preparing cell samples. A better option is to lyse the cells via freeze/thaw cycles or by sonication.